ICoverCalibratorV2CalibratorOn Method

Turns the calibrator on at the specified brightness if the device has calibration capability

Definition

Namespace: ASCOM.DeviceInterface
Assembly: ASCOM.DeviceInterfaces (in ASCOM.DeviceInterfaces.dll) Version: 7.1.3+4851.c5da65c2b
C#
void CalibratorOn(
	int Brightness
)

Parameters

Brightness  Int32
Sets the required calibrator illumination brightness in the range 0 (fully off) to MaxBrightness (fully on).

Remarks

See this link for the canonical definition, which may include further information and implementation requirements: Canonical Definition

Exceptions

InvalidValueExceptionWhen the supplied brightness parameter is outside the range 0 to MaxBrightness.
MethodNotImplementedExceptionWhen CalibratorState returns NotPresent.
NotConnectedExceptionIf the device is not connected.
DriverExceptionAn error occurred that is not described by one of the more specific ASCOM exceptions. Include sufficient detail in the message text to enable the issue to be accurately diagnosed by someone other than yourself.

Revision History

DateVersionDescription
ICoverCalibratorV1Platform 6.5Member added.

See Also